Ficha técnica
Physical dimensions
Dimensions L 151 mm x W 108 mm x H 57 mm
Display 70 x 72 mm, Character height 20 mm
Weight 334 g (without batteries)
Measurement
Technique Oscillometric
NIBP Accuracy validated according to ISO 81060-2 and BHS (A/A Grading)
Systolic Range 60-290 mmHg
Diastolic Range 30-195 mmHg
Pulse Rate Range 30-240 BPM (Beats per min)
Blood Pressure Accuracy ±2% or ±3 mmHg, whichever is greater
Transducer Accuracy ± 3 mmHg over full range in operating conditions
Operating Modes manual
Operating temperature +5°C to +40°C
Operating humidity 15% to 93% relative humidity, non-condensing
Storage temperature -25°C to +70°C
Storage humidity up to 93% relative humidity, non-condensing
Electrical
Operating Voltage 4 – 6 VDC (4 x NiMH or LR6, AA)
Power Consumption Using 5 VDC Nominal
Sleep 60 μW
Idle 105 mW
Measurement 540 mW
Inflation 2.4 W
Storage Capacity 350 Measurements